Catalogue | Optical instruments and Microscopes | Photometers | Polarimeters | Polarimeter SAC-i 589/882 |
InformationDetailsAlternatives/Accessories
Polarimeter SAC-i 589/882
Measurement Scales: Angle of rotation, specific rotation, International Sugar Scale, concentration, purity.
- Wavelength measurement at 589 nm (equivalent to the sodium-D line spectrum) and 882 nm (near infra red)
- Measurement stability in 13 seconds (4 seconds in the high-speed and consecutive measurement mode)
- Resolution of up to 3 decimal places
- Repeatability of ±0.003° Angle of Rotation and ±0.009°Z
- Up to 999 automatic consecutive measurements followed by average value display
- Sleep and timer feature to take measurement later at specified time
- Improved usability due to touch screen technology
- Connectivity to RX Digital Refractometers for automatic purity measurement
Specifications | |
Measurement range Angle of rotation: International standard sugar scale: | -89.999 to +90.000° or -360 to +360° -259 to +259° |
Resolution Angle of rotation: International standard sugar scale: | 0.001° 0.001°Z |
Measurement accuracy Angle of rotation 589 / 882 nm: International standard sugar scale 589 / 882 nm: | ±0.005° / ±0.010° (-45 to +45°) ±0.015°Z / ±0.0700°Z (-130 to +130°Z) |
Measurement temperature range: | 5 to 40°C |
Observation tubes: | 100mm (5ml), 200mm (10ml) |
Power supply: | 100-240V, 50/60Hz a.c. |
Dimensions: Weight: | 600 x 365 x 210mm 20kg |
Polarimeter SAC-i 589/882 |
Type | PK | |
---|---|---|
SAC-i 589/882Prod.No.6.286 467 | 1 |